Create free accountTrade Agreement: NAFTA/Canada FTAs with Peru/Colombia/Panama Tendering Procedures: All interested suppliers may submit a bid Attachment: None Competitive Procurement Strategy: Lowest/Lower Bid Comprehensive Land Claim Agreement: No Nature of Requirements: LOAD FRAME EXTENSOMETER CALIBRATION
W7707-145698/A Gillis, Floyd Telephone No. - (902) 496-5566 Fax No. - (902) 496-5016
Defense Research/Development Canada- Atlantic (DRDC/A) has a requirement for annual calibration (for 3 years) of load frames and extensometers corresponding ASTM standards (Appendix A) by a qualified technician accredited in according to American Association for Laboratory Accreditation A2LA or ISO/IEC 17025 for the purpose of primarily running ASTM E8M tests.
Annual calibration for three ( 3) years of servo-hydraulic and screw driven load frames and accompanying extensometers (listed below) with corresponding ASTM standards.
The following load frames and extensometers equipment shall be calibrated annually for three (3) years:
1 MN Load frame, Instron 1335 SN H0012 · LVDT model 2411-186, SN 3306-64B, range ±150 mm · Load cell model 2518-120, SN UK024, range ±1 MN, 3.5"x12 TPI threads
2. 250 kN Load frame, MTS Landmark 370.25 SN 10313986 · LVDT model 370.25, SN 848604, range ±150 mm · Load cell model 661.21H-01, SN 10306870D, range ±250 kN, m36x2 mm threads
3. 25 kN Load frame, MTS Landmark 370.02 SN 10313996 · LVDT model 370.02, SN 849934, range ±100 mm · Load cell model 661.19H-04, SN 108517A, range ±25 kN, m12x1.25 mm threads · Load cell model 661.19H-02, SN 108517A, range ±10 kN, m12x1.25 mm threads
4. 5 kN Load frame, United SSTM-5KN SN 1209539 · LVDT SN 1209539, range ±1 m · Load cell SN T1I106793, range ±5 kN, 5/8" socket ¼" hitch pins · Built in extensometer model EL-20, SN 5161, range 1 m 5. Static Strain Extensometer · Model 634.12F-24, SN 10304709A, range ±0.5 mm · Model 634.12F-24, SN 10304709C, range ±0.5 mm · Model 634.12F-24, SN 10312611A, range ±4 mm · Model 632.13F-24, SN 10311008C, range ±4 mm · Model 632.13F-20, SN 10310281A, range ±0.15 mm · Model 632.13F-20, SN 10312855, range ±0.15 mm
6. Clip Gauge Extensometers · Model 632.03F-30, SN 10312800, range ±4 mm · Model 632.03F-20, SN 10305679D, range ±12 mm
All listed equipment will be calibrated to: All LVDTs shall be calibrated to ±1.0 % of applied length from 10% - 100% of range, and ± 0.1% of range below 10%
All Load Cells shall be calibrated in compliance with ASTM E4-10 (or newer version)
All Extensometers shall be calibrated in compliance with ASTM E83-10 (or newer version)
All Clip gauges or length devices shall be calibrated in compliance with ASTM E399-10 (or newer version)
All reference standards shall be calibrated in compliance with ASTM E74
All calibration shall be carried out on site at Dockyard Laboratory (Atlantic) located in Bldg D-20, HMC Dockyard, CFB Halifax, Halifax, NS.
Work must be carried out by a qualified technician accredited to American Association for Laboratory Accreditation A2LA or ISO/IEC 17025.
The Request for Proposal (RFP) document forms the solicitation requirement.
This office provides service to the public in English. Delivery Date: Above-mentioned
The Crown retains the right to negotiate with suppliers on any procurement.
Documents may be submitted in either official language of Canada.
Not stated in the notice: procurement type, procurement method, selection criteria, region of delivery, region of opportunity, expected contract period, end user